Electrosurgery
A surgeon using a monopolar RF electrosurgical instrument to coagulate (and desiccate) tissue in the excision of a lipoma
MeSHD004598
[edit on Wikidata]

Electrosurgery is the application of a high-frequency (radio frequency) alternating polarity, electrical current to biological tissue as a means to cut, coagulate, desiccate, or fulgurate tissue.[1][2][3][4][5][6][7] (These terms are used in specific ways for this methodology—see below.) Its benefits include the ability to make precise cuts with limited blood loss. Electrosurgical devices are frequently used during surgical operations helping to prevent blood loss in hospital operating rooms or in outpatient procedures.[8]

In electrosurgical procedures, the tissue is heated by an electric current. Although electrical devices that create a heated probe may be used for the cauterization of tissue in some applications, electrosurgery refers to a different method than electrocautery. Electrocautery uses heat conduction from a probe heated to a high temperature by a direct electrical current (much in the manner of a soldering iron). This may be accomplished by direct current from dry-cells in a penlight-type device.

Electrosurgery, by contrast, uses radio frequency (RF) alternating current to heat the tissue by RF induced intracellular oscillation of ionized molecules that result in an elevation of intracellular temperature. When the intracellular temperature reaches 60 degrees C, instantaneous cell death occurs. If tissue is heated to 60–99 degrees C, the simultaneous processes of tissue desiccation (dehydration) and protein coagulation occur. If the intracellular temperature rapidly reaches 100 degrees C, the intracellular contents undergo a liquid to gas conversion, massive volumetric expansion, and resulting explosive vaporization.

Appropriately applied with electrosurgical forceps, desiccation and coagulation result in the occlusion of blood vessels and halting of bleeding. While the process is technically a process of electrocoagulation, the term "electrocautery" is sometimes loosely, nontechnically and incorrectly used to describe it. The process of vaporization can be used to ablate tissue targets, or, by linear extension, used to transect or cut tissue. While the processes of vaporization/ cutting and desiccation/coagulation are best accomplished with relatively low voltage, continuous or near continuous waveforms, the process of fulguration is performed with relatively high voltage modulated waveforms. Fulguration is a superficial type of coagulation, typically created by arcing modulated high voltage current to tissue that is rapidly desiccated and coagulated. The continued application of current to this high impedance tissue results in resistive heating and the achievement of very high temperatures—enough to cause breakdown of the organic molecules to sugars and even carbon, thus the dark textures from carbonization of tissue.

Diathermy is used by some as a synonym for electrosurgery but in other contexts diathermy means dielectric heating, produced by rotation of molecular dipoles in a high frequency electromagnetic field. This effect is most widely used in microwave ovens or some tissue ablative devices which operate at gigahertz frequencies. Lower frequencies, allowing for deeper penetration, are used in industrial processes.

RF electrosurgery is commonly used in virtually all surgical disciplines including dermatological, gynecological, cardiac, plastic, ocular, spine, ENT, maxillofacial, orthopedic, urological, neuro- and general surgical procedures as well as certain dental procedures.

RF electrosurgery is performed using a RF electrosurgical generator (also referred to as an electrosurgical unit or ESU) and a handpiece including one or two electrodes—a monopolar or bipolar instrument. All RF electrosurgery is bipolar so the difference between monopolar and bipolar instruments is that monopolar instruments comprise only one electrode while bipolar instruments include both electrodes in their design.

The monopolar instrument called an "active electrode" when energized, requires the application of another monopolar instrument called a "dispersive electrode" elsewhere on the patient's body that functions to 'defocus' or disperse the RF current thereby preventing thermal injury to the underlying tissue. This dispersive electrode is frequently and mistakenly called a "ground pad" or "neutral electrode". However virtually all currently available RF electrosurgical systems are designed to function with isolated circuits—the dispersive electrode is directly attached to the ESU, not to "ground". The same electrical current is transmitted across both the dispersive electrode and the active electrode, so it is not "neutral". The term "return electrode" is also technically incorrect since alternating electrical currents refer to alternating polarity, a circumstance that results in bidirectional flow across both electrodes in the circuit.

Bipolar instruments generally are designed with two "active" electrodes, such as a forceps for sealing blood vessels. However, the bipolar instrument can be designed such that one electrode is dispersive. The main advantage of bipolar instruments is that the only part of the patient included in the circuit is that which is between the two electrodes, a circumstance that eliminates the risk of current diversion and related adverse events. However, except for those devices designed to function in fluid, it is difficult to vaporize or cut tissue with bipolar instruments.

  1. ^ Hainer BL, "Fundamentals of electrosurgery", Journal of the American Board of Family Practice, 4(6):419–26, 1991 Nov.–Dec.
  2. ^ Electrosurgery for the Skin Archived 2008-05-17 at the Wayback Machine, Barry L. Hainer M.D., Richard B. Usatine, M.D., American Family Physician (Journal of the American Academy of Family Physicians), 2002 Oct 1;66(7):1259–66.
  3. ^ "A Simple Guide to the Hyfrecator 2000" Archived 2007-09-28 at the Wayback Machine. Schuco International (London) Ltd.
  4. ^ Boughton RS, Spencer SK (Apr 1987). "Electrosurgical fundamentals". J Am Acad Dermatol. 16 (4): 862–7. doi:10.1016/s0190-9622(87)70113-3. PMID 3571547.
  5. ^ Bouchier G, "The fundamentals of electro-surgery. High frequency current generators", Cah Prothese, 1980 Jan;8(29):95–106. In French.
  6. ^ Oringer MJ (Jan 1960). "Fundamentals of electrosurgery". J Oral Surg Anesth Hosp Dent Serv. 18: 39–49. PMID 14429020.
  7. ^ Reidenbach HD (Apr 1993). "Fundamentals of bipolar high-frequency surgery". Endosc Surg Allied Technol. 1 (2): 85–90. PMID 8055306.
  8. ^ Cite error: The named reference McCauley2003 was invoked but never defined (see the help page).

and 29 Related for: Electrosurgery information

Request time (Page generated in 0.5758 seconds.)

Electrosurgery

Last Update:

Electrosurgery is the application of a high-frequency (radio frequency) alternating polarity, electrical current to biological tissue as a means to cut...

Word Count : 3688

Gingivectomy

Last Update:

gradually increase if the patients’ plaque control is up to standard. Electrosurgery is defined as the intentional passage of high-frequency waveforms or...

Word Count : 2880

Hyfrecator

Last Update:

A hyfrecator is a low-powered medical apparatus used in electrosurgery on conscious patients, usually in an office setting. It is used to destroy tissue...

Word Count : 1273

Diathermy

Last Update:

better known as "electrosurgery". (It is also referred to occasionally as "electrocautery", but see disambiguation below.) Electrosurgery and surgical diathermy...

Word Count : 2604

Harmonic scalpel

Last Update:

that (unlike electrosurgery) uses ultrasonic vibrations to cut and cauterize tissue. The harmonic scalpel may be superior to electrosurgery in that it can...

Word Count : 304

Fibrous papule of the nose

Last Update:

necessary. Treatments for cosmetic reasons include shave excision and electrosurgery. Following treatment, recurrence is rare. Skin lesion James, William...

Word Count : 265

Radio frequency

Last Update:

years, and now include diathermy, hyperthermy treatment of cancer, electrosurgery scalpels used to cut and cauterize in operations, and radiofrequency...

Word Count : 1091

Formaldehyde

Last Update:

suppress. In operating rooms, formaldehyde is produced as a byproduct of electrosurgery and is present in surgical smoke, exposing surgeons and healthcare workers...

Word Count : 8177

Medical applications of radio frequency

Last Update:

years, and now include diathermy, hyperthermy treatment of cancer, electrosurgery scalpels used to cut and cauterize in operations, and radiofrequency...

Word Count : 921

Biophysics

Last Update:

development in the mid-20th century. He was a leader in developing electrosurgery. The popularity of the field rose when the book What Is Life? by Erwin...

Word Count : 1649

Surgical instrument

Last Update:

calipers Energy systems tools are used to cut tissues or seal vessels Electrosurgery is a modern technology that uses high frequency electrical current to...

Word Count : 2533

Abdominal surgery

Last Update:

low-certainty evidence that there is no difference between using scalpel and electrosurgery in infection rates during major abdominal surgeries. Wikisource has...

Word Count : 969

Lipoma

Last Update:

potential methods based on tissue-targeted heating include cauterization, electrosurgery, and harmonic scalpel. Patients who have a lipoma removed are usually...

Word Count : 3030

Liver spot

Last Update:

unsightly and some people choose to have them removed. This can be done by electrosurgery, laser treatment, cryotherapy, or the use of depigmentation agents,...

Word Count : 981

Actinic cheilitis

Last Update:

5-fluorouracil, imiquimod, scalpel vermillionectomy, chemical peel, electrosurgery, and carbon dioxide laser vaporization. These curative treatments attempt...

Word Count : 1455

Edwin Beer

Last Update:

treatment of papillary bladder tumors. He is recognized as the founder of electrosurgery of the bladder. Beer was born and raised in New York City. His parents...

Word Count : 559

Gum lift

Last Update:

would need to be obtained privately, where the charge is variable. Electrosurgery is defined as “the intentional passage of high frequency waveforms or...

Word Count : 1980

Cauterization

Last Update:

Covidien (no longer part of Tyco Healthcare), explaining the basics of electrosurgery Examples of Cauterizing the Wound in Cinema - Daily Film Dose Villafane...

Word Count : 2159

Surgical oncology

Last Update:

times. The most effective surgical oncology techniques are: cryosurgery electrosurgery laparoscopic surgery laser surgery mohs surgery radiofrequency ablation...

Word Count : 755

Esu

Last Update:

charge Evolutionarily significant unit Electrosurgical unit, used in electrosurgery Search for "esu"  or "E-S-U" on Wikipedia. Esus (disambiguation) All...

Word Count : 206

Cryosurgery

Last Update:

away within a few days. Chemical cautery Cryoneurolysis Cryotherapy Electrosurgery "Cryotherapy - DermNet New Zealand". dermnetnz.org. "Cryotherapy: Overview...

Word Count : 1098

Dental instrument

Last Update:

bleeding following soft tissue procedures, and some risks of alternative electrosurgery procedures are avoided. A dental torque wrench or restorative torque...

Word Count : 1644

Toluene

Last Update:

occurs as an indoor air pollutant in a number of processes including electrosurgery, and can be removed from the air with an activated carbon filter. Similarly...

Word Count : 3184

Surgery

Last Update:

low-temperature cryoablation to freeze and destroy a target tissue. Electrosurgery involves use of electrocautery to cut and coagulate tissue. Microsurgery...

Word Count : 11791

List of instruments used in ophthalmology

Last Update:

etc. Electrolysis used for permanent hair removal Electrocautery for electrosurgery Phacoemulsification used for extraction of a cataract affected lens...

Word Count : 278

Argon

Last Update:

procedure called "argon-enhanced coagulation", a form of argon plasma beam electrosurgery. The procedure carries a risk of producing gas embolism and has resulted...

Word Count : 4633

Electrical injury

Last Update:

not stimulate the nerves strongly enough to cause cardiac arrest (see electrosurgery). Also important is the pathway: if the current passes through the chest...

Word Count : 5935

Blunt dissection

Last Update:

dissection, the practice of slicing through tissues with scalpels, scissors, electrosurgery, or other advanced technologies usually employing heat. New devices...

Word Count : 197

Surgical smoke

Last Update:

Surgical smoke is the by-product produced by electrosurgery, laser tissue ablation, or other surgical techniques. Surgical smoke, as a health threat to...

Word Count : 877

PDF Search Engine © AllGlobal.net