Global Information Lookup Global Information

Photoplethysmogram information


Photoplethysmography
Representative PPG taken from an ear pulse oximeter. Variation in amplitude are from Respiratory Induced Variation.
MeSHD017156

A photoplethysmogram (PPG) is an optically obtained plethysmogram that can be used to detect blood volume changes in the microvascular bed of tissue.[1][2] A PPG is often obtained by using a pulse oximeter which illuminates the skin and measures changes in light absorption.[3] A conventional pulse oximeter monitors the perfusion of blood to the dermis and subcutaneous tissue of the skin.

Finger pulse oximeter

With each cardiac cycle the heart pumps blood to the periphery. Even though this pressure pulse is somewhat damped by the time it reaches the skin, it is enough to distend the arteries and arterioles in the subcutaneous tissue. If the pulse oximeter is attached without compressing the skin, a pressure pulse can also be seen from the venous plexus, as a small secondary peak.

The change in volume caused by the pressure pulse is detected by illuminating the skin with the light from a light-emitting diode (LED) and then measuring the amount of light either transmitted or reflected to a photodiode.[4] Each cardiac cycle appears as a peak, as seen in the figure. Because blood flow to the skin can be modulated by multiple other physiological systems, the PPG can also be used to monitor breathing, hypovolemia, and other circulatory conditions.[5] Additionally, the shape of the PPG waveform differs from subject to subject, and varies with the location and manner in which the pulse oximeter is attached.

Although PPG sensors are in common use in a number of commercial and clinical applications, the exact mechanisms determining the shape of the PPG waveform are not yet fully understood.[6]

  1. ^ Allen J (March 2007). "Photoplethysmography and its application in clinical physiological measurement". Physiological Measurement. 28 (3): R1–39. doi:10.1088/0967-3334/28/3/R01. PMID 17322588. S2CID 11450080.
  2. ^ Kyriacou PA, Allen J, eds. (2021). Photoplethysmography: Technology, Signal Analysis and Applications. Elsevier.
  3. ^ Shelley K, Shelley S, Lake C (2001). "Pulse Oximeter Waveform: Photoelectric Plethysmography". In Lake C, Hines R, Blitt C (eds.). Clinical Monitoring. W.B. Saunders Company. pp. 420–428.
  4. ^ Pelaez EA, Villegas ER (2007). "LED power reduction trade-offs for ambulatory pulse oximetry". 2007 29th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. Vol. 2007. pp. 2296–2299. doi:10.1109/IEMBS.2007.4352784. ISBN 978-1-4244-0787-3. PMID 18002450. S2CID 34626885.
  5. ^ Reisner A, Shaltis PA, McCombie D, Asada HH (May 2008). "Utility of the photoplethysmogram in circulatory monitoring". Anesthesiology. 108 (5): 950–958. doi:10.1097/ALN.0b013e31816c89e1. PMID 18431132.
  6. ^ Charlton PH, Kyriaco PA, Mant J, Marozas V, Chowienczyk P, Alastruey J (March 2022). "Wearable Photoplethysmography for Cardiovascular Monitoring". Proceedings of the IEEE. Institute of Electrical and Electronics Engineers. 110 (3): 355–381. doi:10.1109/JPROC.2022.3149785. PMC 7612541. PMID 35356509.

and 11 Related for: Photoplethysmogram information

Request time (Page generated in 0.5483 seconds.)

Photoplethysmogram

Last Update:

A photoplethysmogram (PPG) is an optically obtained plethysmogram that can be used to detect blood volume changes in the microvascular bed of tissue. A...

Word Count : 2131

Respiratory rate

Last Update:

such as estimation of respiratory rate from the electrocardiogram, photoplethysmogram, or accelerometry signals. Breathing rate is often interchanged with...

Word Count : 1183

Photoplethysmogram variability

Last Update:

The photoplethysmogram (PPG) measurement made at a peripheral site, such as the finger, ear or forehead represents the volume of blood in the vessel at...

Word Count : 596

Ballistocardiography

Last Update:

Nongjian (2017). "Simultaneous Monitoring of Ballistocardiogram and Photoplethysmogram Using a Camera". IEEE Transactions on Biomedical Engineering. 64 (5):...

Word Count : 548

Laser Doppler imaging

Last Update:

of blood flow measured by laser Doppler holography in the digit (photoplethysmogram) and the eye fundus has a pulse-shaped profile with time. These remote...

Word Count : 1028

Physiological Signal Based Security

Last Update:

Body Area Networks (BANs) constantly interact with their physical environment with the help of sensors. Sensors collect process as well as communicate...

Word Count : 646

Pulse oximetry

Last Update:

blood sample) and changes in blood volume in the skin, producing a photoplethysmogram that may be further processed into other measurements. The pulse oximeter...

Word Count : 6979

Hemodynamics

Last Update:

Esophogeal doppler Hemodynamics of the aorta Impedance cardiography Photoplethysmogram Laser Doppler imaging Windkessel effect Functional near-infrared spectroscopy...

Word Count : 6415

Samsung Bio Processor

Last Update:

integrates five AFEs including bioelectrical impedance analysis (BIA), photoplethysmogram (PPG), electrocardiogram (ECG), skin temperature, and galvanic skin...

Word Count : 235

Holographic interferometry

Last Update:

microvascular blood flow imaging, and wide-field measurement of photoplethysmograms by detection of out-of-plane tissue motion. The wide temporal bandwidth...

Word Count : 1567

Ronald Aarts

Last Update:

respiratory rate, can be performed without hindering the patient using a photoplethysmogram (PPG). A PPG sensor can easily be built into a bracelet such as a...

Word Count : 1224

PDF Search Engine © AllGlobal.net